Research Facilities
Jasco FT-IR-6300 Fourier Transform Infrared Spectrometer
- Single beam optical system with high intensity ceramic light source;
- Wavenumber range: 7800 – 150 cm-1;
- Wavenumber resolution / accuracy: 0.07 cm-1 / ± 0.01 cm-1;
- Interferometer system: 28° Michelson interferometer with auto-alignment mechanism, sealed structure, and gold coated mirrors;
- Secure-lock beam splitter catch system with auto-alignment mechanism;
- Available Beam Splitters: Ge/KBr (7800 – 400 cm-1) and Mylar-5 (600 – 150 cm-1);
- KRS-5 and PE interchangeable windows of the sample compartment;
- Peltier temperature control of the detector system with automatic detector exchange mechanism;
- Available detectors: DLATGS (7800 – 360 cm-1) and DTGS(PE) (450 – 150 cm-1);
- Up to two more external detectors may be installed via the external detector port.
Accessories:
1. Pike MIRacle – universal ATR sampling accessory
- Available crystals: Si (7800 – 1500 cm-1, 475-150 cm-1)
Si/ZnSe (7800 – 550 cm-1)
2. Pike Mid-IR IntegratIR – 3 inch diameter Au coated integrating sphere used for the measurement of absolute and relative diffuse reflectance
- LN2-cooled MCT detector;
- Wavenumber range: 7800 – 550 cm-1.